MAUDE Adverse Event Report: OBERDORF SYNTHES PRODUKTIONS GMBH 1.5MM TI CORTEX SCR SLF-TPNG WITH T4 STARDRIVE RECESS 10MM; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OBERDORF SYNTHES PRODUKTIONS GMBH 1.5MM TI CORTEX SCR SLF-TPNG WITH T4 STARDRIVE RECESS 10MM; SCREW, FIXATION, BONE Back to Search Results
Model Number 04.214.110
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 06/13/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient case number reported as (b)(6).Additional procode: hrs.Lot number is unknown.Device malfunctioned intra-operatively and was not implanted / explanted.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Reporter phone number and email address are unknown.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes europe reports an event in (b)(6) as follows: during an implant placement procedure on (b)(6) 2018, the heads of two titanium (ti) cortex screws broke due to the worn tip of two (2) stardrive screwdriver shafts.All device fragments were easily removed and new screws were implanted to the patient.The procedure was completed successfully with a reported delay of forty five (45) minutes.There is no other reported patient consequence.This report is for a 1.5mm ti cortex scr slf-tpng with t4 stardrive recess 10mm.This is report 4 of 4 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.5MM TI CORTEX SCR SLF-TPNG WITH T4 STARDRIVE RECESS 10MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
OBERDORF SYNTHES PRODUKTIONS GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key7681273
Report Number8030965-2018-54959
Device Sequence Number1452655
Product Code HWC
UDI-Device Identifier10886982120854
UDI-Public(01)10886982120854
Combination Product (Y/N)N
Initial Reporter CountryCS
PMA/510(K) Number
K112583
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number04.214.110
Device Catalogue Number04.214.110
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/14/2018
Initial Report FDA Received Date07/12/2018
Patient Sequence Number1
-
-