| Catalog Number |
NTSE-030115-UDH |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/27/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Pma/510(k) # - exempt.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
|
| |
|
Event or Problem Description
|
|
It was reported before the procedure, the physician wanted to use the ncircle tipless stone extractor but he found the basket wire was broken.Thus, he used a new ncircle tipless stone extractor to complete the ureteroscopy and laser stone fragmentation (ursl) procedure.There were no adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation - evaluation: visual inspection and functional testing of the returned device was conducted during the investigation.A review of complaint history, the device history record, instructions for use, manufacturing instructions, quality control data, and specifications was also performed.One device was returned for investigation.The device was returned with the handle and basket formation in the closed positions.The collet knob is tight and secure.The male luer lock adapter (mlla) is finger tight.The polyethylene terephthalate tubing (pett) measures 3 cm in length.A visual examination noted kinks in the basket sheath located at 37 cm from the distal tip, again at 56.5 cm from the distal tip, and 78.7 cm from the distal tip.The basket formation was noted to have a broken wire 2 mm from the cannula.A functional test noted the handle does not actuate the basket formation.The handle was disassembled.The basket sheath and support sheath are still adhered.The basket formation can be manually actuated.The device history records were reviewed and found no non-conformances that would have contributed to the reported failure mode.A review of complaint history shows this complaint is the only one associated with lot number 8544303.The instructions for use (ifu) provides the following precautions: enclose the device in the sheath before removing from the tray/holder.Do not use excessive force to manipulate this device.Damage to the device may occur.A review of relevant manufacturing documents was conducted.It was concluded that the device aspect in question was visually/functionally inspected by quality control and no related gaps in production or processing controls were noted.Devices are inspected for damage and functionality prior to packaging.The returned device was reported to have a broken basket wire, with the wire broken near the basket cannula.A cause for the broken wire could not be determined.No damage to the distal end of the device was found that could have damaged the basket wire.It was also found that the basket did not function and the sheath was kinked in multiple locations along its length.The non-functional basket was not mentioned by the customer.No conclusion could be drawn based on the information provided and the evaluation findings.Per the quality engineering risk assessment, no further action is required.Cook medical has notified the appropriate personnel and will continue to monitor this device via the complaints database for similar complaints.
|
| |
|
Event or Problem Description
|
|
It was reported before the procedure, the physician wanted to use the ncircle tipless stone extractor but he found the basket wire was broken.Thus, he used a new ncircle tipless stone extractor to complete the ureteroscopy and laser stone fragmentation (ursl) procedure.There were no adverse consequences to the patient.
|
| |
|