• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH MNTR LS SCREW 3.5X30; ORTHOSIS, CERVICAL PEDICLE SCREW SPINAL FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH MNTR LS SCREW 3.5X30; ORTHOSIS, CERVICAL PEDICLE SCREW SPINAL FIXATION Back to Search Results
Model Number 188320330
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #: (b)(4).Visual examination of the mntr ls screw reveals signs of operative use as evidenced by superficial markings, with damage to the drive feature of the screw head and the inner cap (saddle) is dislodged and fell apart from its intended location.Exerting a significant amount of force on the tulip head and saddle, potentially at a significant angle, may be sufficient to dislodge the saddle.A dhr review did not identify any issues that could have cause or contributed to the reported event.All complaint trends will be evaluated as a part of the depuy spine monthly complaint review meeting.The root cause of the mntr ls screw falling apart cannot be determined from the sample and the information provided.A potential root cause may be excessive force inadvertently placed on the drive feature of the screw during tightening, potentially at a significant angle, resulting in the saddle falling apart during use.As no issues were identified in the manufacturing and release of this device that could have contributed to the problem reported by the customer, this complaint will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
When the surgeon was doing final tightening with torque drivers the locking screw and the nut jumped out.Patient consequence?: no.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MNTR LS SCREW 3.5X30
Common Device Name
ORTHOSIS, CERVICAL PEDICLE SCREW SPINAL FIXATION
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key7681339
Report Number1526439-2018-50643
Device Sequence Number16302237
Product Code NKG
UDI-Device Identifier10705034158136
UDI-Public(01)10705034158136
Combination Product (Y/N)N
PMA/510(K) Number
K041203
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Caregivers
Type of Report Initial
Report Date (Section B) 02/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number188320330
Device Catalogue Number188320330
Device Lot NumberATFFG2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2018
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/13/2018
Initial Report FDA Received Date07/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
-
-