| Model Number |
693558 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Fracture (1260); Impedance Problem (2950)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Information (3190)
|
| Date of Event |
06/16/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant product(s): dtba1d1 crt-d, implanted: (b)(6) 2017.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient¿s right ventricular (rv) lead triggered a lead integrity alert (lia) as well as a lead noise warning for oversensing-noise episodes and lead impedance warnings for rising pacing impedance.Oversensing was also noted upon manipulation of the patient¿s left arm.A fracture of the rv lead was confirmed.The rv lead was capped and replaced.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|