• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI SPRINT QUATTRO SECURE S; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI SPRINT QUATTRO SECURE S; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER Back to Search Results
Model Number 693558
Medical Device Problem Codes Signal Artifact/Noise (1036); Fracture (1260); Impedance Problem (2950)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Information (3190)
Date of Event 06/16/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant product(s): dtba1d1 crt-d, implanted: (b)(6) 2017.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient¿s right ventricular (rv) lead triggered a lead integrity alert (lia) as well as a lead noise warning for oversensing-noise episodes and lead impedance warnings for rising pacing impedance.Oversensing was also noted upon manipulation of the patient¿s left arm.A fracture of the rv lead was confirmed.The rv lead was capped and replaced.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPRINT QUATTRO SECURE S
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key7681881
Report Number2649622-2018-11317
Device Sequence Number19174880
Product Code LWS
UDI-Device Identifier00643169535909
UDI-Public00643169535909
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P920015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/13/2017
Device Model Number693558
Device Catalogue Number693558
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/19/2018
Supplement Date Received by Manufacturer08/08/2018
Initial Report FDA Received Date07/12/2018
Supplement Report FDA Received Date10/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
419478 LEAD, 5076-52 LEAD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age79 YR
Patient Weight88
-
-