| Model Number |
T066-0205 |
| Medical Device Problem Codes |
Break (1069); Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The complainant reported to the company that a guided angled awl tip broke inside a cage during surgery.A dhr review was performed and there were no manufacturing anomalies identified.The device met all required specifications prior to being released for distribution.The damage to the distal tip of the awl could be replicated if lateral force was applied while seated in an implant, or from failure to use the guided inserter.
|
| |
|
Event or Problem Description
|
|
The complainant reported to the company that a guided angled awl tip broke inside a cage during surgery.There were no known patient complications.
|
| |
|