• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TI DUR REG FLUTED STEM 23X80MM; KNEE JOINT PATELLOFEMOROTIBIAL POLYMER/METAL/POLYMER SEMI-CONSTRAINED CEMENTED P

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TI DUR REG FLUTED STEM 23X80MM; KNEE JOINT PATELLOFEMOROTIBIAL POLYMER/METAL/POLYMER SEMI-CONSTRAINED CEMENTED P Back to Search Results
Catalog Number 64786665
Medical Device Problem Code Osseointegration Problem (3003)
Health Effect - Clinical Code Injury (2348)
Date of Event 06/14/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A supplemental report will be submitted upon completion of the investigation.
 
Event or Problem Description
After implant was removed, it was noticed that the stem was lose and the was a crack on the connection between stem and femur component.
 
Event or Problem Description
After implant was removed, it was noticed that the stem was lose and the was a crack on the connection between stem and femur component.
 
Additional Manufacturer Narrative
An event regarding revision due to loosening involving a duracon stem was reported.The event was not confirmed.Method & results: product evaluation and results: visual inspection: visual inspection of the returned devices was performed.The stem extender component was returned assembled to the mrh femoral component.The stem extender was unscrewed from the mrh femoral component as part of the evaluation.The mrh femoral component is consistent with an explanted device.There are nicks and scratches on the component, most likely from in vivo use and subsequent explantation.There is no obvious damage to the component.There is no damage evident on the mrh femoral boss/thread.Dimensional inspection was not performed as the dimensional aspects are not in question- based on the manufacturing date of the device it would have been implanted in this very active patient for at least 10 years.Medical records received and evaluation: not performed as no medical records were received for review.Product history review: indicated all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusion: the stem extender component was returned assembled to the mrh femoral component.The stem extender was unscrewed from the mrh femoral component as part of the evaluation.The mrh femoral component is consistent with an explanted device.There are nicks and scratches on the component, most likely from in vivo use and subsequent explantation.There is no obvious damage to the component.There is no damage evident on the mrh femoral boss/thread.Dimensional inspection was not performed as the dimensional aspects are not in question- based on the manufacturing date of the device it would have been implanted in this very active patient for at least 10 years.The exact cause of the event could not be determined because insufficient information was provided.Additional information, including operative reports, progress notes, patient history and x-rays are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TI DUR REG FLUTED STEM 23X80MM
Common Device Name
KNEE JOINT PATELLOFEMOROTIBIAL POLYMER/METAL/POLYMER SEMI-CONSTRAINED CEMENTED P
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7682072
Report Number0002249697-2018-02149
Device Sequence Number848073
Product Code JWH
UDI-Device Identifier07613327044973
UDI-Public07613327044973
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K973164
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number64786665
Device Lot NumberTBYC91A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/26/2018
Initial Date Received by Manufacturer 06/14/2018
Supplement Date Received by Manufacturer07/26/2018
Initial Report FDA Received Date07/12/2018
Supplement Report FDA Received Date08/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-