| Model Number |
ASPC28-3PCE. |
| Medical Device Problem Code |
Disconnection (1171)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/06/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The complaint and returned sample were forwarded to the contract device manufacturer for evaluation.
|
| |
|
Event or Problem Description
|
|
The patient took out the catheter inserted the day before.The suture wing and the suture were still attached to the patient's skin, the suture was not ripped off.The catheter separated from the suture wing.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the investigation performed and the results of the pull-test inspection, a root cause could not be attributable to manufacture process.We are unable to determine the exact cause of the event; however, it appears that at some point enough force was applied to the catheter to dislodge it from the suture wing.
|
| |
|