| Model Number |
977D260 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received from a manufacturer representative (rep) regarding a trial patient being implanted with an unknown spinal cord stimulator model.It was reported that on (b)(6) 2018 during a trial lead implant procedure, the trial lead was damaged; an electrode looked like it had broken off of the lead when the doctor was deploying the lead from the needle/introducer.The doctor decided to use another lead.No symptoms were reported.No further complications were reported or anticipated.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis results were not available at the time of this report.A follow-up report will be sent when analysis is completed.Device conclusion code does not apply.Device result code does not apply.Device method code.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information was received from a manufacturer representative regarding the patient.It was reported that the physician deployed the lead and he rotated the needle.It appeared that the top electrode had bent back and almost looked as if it had broken off.The lead was removed.There were no patient symptoms or complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis of the lead (b)(4).Found that the tip of the distal end was bent.Product analysis (b)(4):analysis information -- 2018-09-14 11:35:11 cst pli# 10 product id# 977d260 below is unedited, system generated text based on the analysis finding code(s) and test results.Analysis identified the {x} electrode at the distal end of the lead was bent.Electrical testing of the lead determined continuity was complete and no electrical shorts were identified between the circuits.Analysis observed the distal end of the lead was bent.{the lead and stylet wire are both severely bent on the distal side of the #1 electrode.} product id 3550-32 product type accessory product id neu_stylet_acc product type accessory.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|