• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI) CARESTATION 650; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI) CARESTATION 650; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Suction Problem (2170)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/14/2018
Event Type  malfunction  
Manufacturer Narrative
The customer biomedical engineer troubleshot the issue with ge healthcare technical support.The kinked tubing in the rear of the unit was re-routed to resolve the reported issue.No report of patient involvement.Unique device identifier - (b)(4).The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.
 
Event Description
The hospital reported inadequate suction.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARESTATION 650
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI)
no. 19 changjiang road
national hi-tech dev. zone
wuxi, 21402 8
CH  214028
Manufacturer (Section G)
GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI)
no. 19 changjiang road
national hi-tech dev. zone
wuxi, 21402 8
CH   214028
Manufacturer Contact
john szalinski
3000 n grandview blvd.
waukesha, WI 53188
8479711808
MDR Report Key7683637
MDR Text Key113882199
Report Number9710602-2018-00179
Device Sequence Number1
Product Code BSZ
UDI-Device Identifier00840682103947
UDI-Public01008406821039470
Combination Product (y/n)N
PMA/PMN Number
K151570
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial
Report Date 07/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 06/14/2018
Initial Date FDA Received07/12/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/10/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-