Catalog Number 354220
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device involved has been received for evaluation and the investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event or Problem Description
As reported by the user facility: event 3: tubing separating from distal y-site during infusion of solution.Veterinary technicians noticed that the tubing that connects into the housing of the distal y-site was separating just by pulling lightly.The staff came back to a patient and noticed the set had separated even without manual manipulation.This has happened on 5 occasions now, but the sets have been thrown out.The staff were checking the strength of each set before attaching to a patient to avoid any issues.It appears the set has been exposed to some blood from an animal.No injury reported.
 
Additional Manufacturer Narrative
Event 3: this report has been identified as b.Braun medical internal report number (b)(4).One (1) used sample without packaging was returned for evaluation.The returned sample was tested for leakage per specification with passing results.All bonded joints on the set were also pull tested per specification with passing results.The defect of disconnection of a bonded joint was not confirmed.Review of the discrepancy management system database was unable to be performed because the lot number was reported as unknown.We will maintain this report for further references and continue to monitor other reports for similar occurrences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARESITE®
Common Device Name
SET, ADMINISTRATION, INTRA
Manufacturer (Section D)
B. BRAUN DOMINICAN REPUBLIC INC.
las americas industrial park
km22 autopista las americas
santo domingo,
DR 
MDR Report Key7683825
Report Number9614279-2018-00112
Device Sequence Number13802374
Product Code FPA
UDI-Device Identifier04046964182358
UDI-Public04046964182358
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K942988
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number354220
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/26/2018
Initial Date Received by Manufacturer 06/18/2018
Supplement Date Received by Manufacturer06/18/2018
Initial Report FDA Received Date07/12/2018
Supplement Report FDA Received Date08/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-