| Model Number |
AC*T DIFF 2 |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
Test Result (2695)
|
| Date of Event |
06/15/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
On 18-jun-2018 the field service engineer (fse) was at the customer site and observed that the coulter ac·t diff 2 analyzer was generating high platelet results with electrical noise on the plt histogram.The fse replaced the vacuum isolation chamber (vic) which eliminated the electrical crosstalk and resolved the noise problem.Service activity performed was verified to meet the specified requirements per established procedures.Results meet published performance specifications.The cause of the high plt on qc and patient samples was a defective vacuum isolator chamber.Patient information was not provided by the customer.Bec internal identifier - (b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reported having an ongoing issue with erroneously high platelet (plt) for qc and patient samples on their coulter ac·t diff 2 analyzer.Erroneous results were not reported outside of the laboratory.There was no change to patient treatment as a result of this event.Patient and qc data was not provided by the customer.
|
| |
|