| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/18/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.An evaluation is in process.
|
| |
|
Event or Problem Description
|
|
The customer observed falsely depressed results while using total bilirubin reagents.The following data was provided.Sid (b)(6) initial and repeat results read 0.481 mg/dl (reported as less than 1.50 mg/dl).Repeat using another analyzer 2.67, 2.74 mg/dl.No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation of the customer issue included a complaint text review, a search for similar complaints, device history review, instrument log review, and labeling review.Returns were not available.No adverse trend was identified for the customer issue.Device history review did not identify any issues that may have caused the customer issue.Log review did not identify any issues that contributed to the customer issue.Labeling was reviewed and found to be adequate.Based on all available information and abbott diagnostics evaluation, no systemic issue was identified and no product deficiency was found.
|
| |
|