| Model Number |
ZXT150 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/07/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If implanted, give date: not applicable, as lens was removed/replaced in the initial surgery.If explanted, give date: not applicable, as lens was removed/replaced in the initial surgery.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that intraocular lens had torn/broken haptic.On performing multiple follow up attempts it was not confirmed if only haptic was torn of optic was torn.The event was observed while inserting into the eye.The lens was removed and replaced in same surgical procedure.Customer confirmed that there was no patient injury, no incision enlargement, no surgical or medical intervention.Patient was fine at discharge.No further information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation: yes.Returned to manufacturer on: 8/22/2018.Device returned to manufacturer: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.The lens was inspected by a qualified final inspection operator using 12x magnification.It can be seen that the lens is contaminated.This is expected as the lens was transported from controlled environment during manufacturing to a non-sterile, non-environmental controlled area.The was no damage seen on the lens.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|