• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ULTRASCORE FOCUSED FORCE PTA BALLOON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ULTRASCORE FOCUSED FORCE PTA BALLOON Back to Search Results
Model Number US35130515
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/13/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.Expiry date: 03/2020; manufacturing date: 04/2018.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported through the results of a clinical trial that during an angioplasty procedure with a focused force percutaneous transluminal angioplasty balloon dilation catheter, upon removal of the device, a wire was noted to be detached from the tip of the balloon.There was no reported impact or consequence to the patient.
 
Additional Manufacturer Narrative
Manufacturing review: the lot number has been provided and the lot device history records have been reviewed.The lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported for this lot number and issue to date.Investigation summary: the device was returned for evaluation.A visual inspection found a break in one of the scoring wires at the distal tip wire coil.The wire had pulled out of the sleeve.The wire was not completely detached from the device as it was still attached at the proximal end of the wire.Therefore, the investigation is confirmed for a break in the scoring wire.The distal end of the wire at the break was measured and found to be within the acceptable specification for the wire diameter.Additionally, the distal end of the device is 100% inspected to verify the bond between the wire and laminate.Also each lot is inspected at a predetermined sample size for a visual inspection of the tip.Based on the feedback from the user and the returned sample condition, the definitive root cause for the break could not be determined.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.(manufacturing date: 04/2018).
 
Event or Problem Description
It was reported through the results of a clinical trial that during an angioplasty procedure with a focused force percutaneous transluminal angioplasty balloon dilation catheter, upon removal of the device, a wire was noted to be detached from the tip of the balloon.There was no reported impact or consequence to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASCORE FOCUSED FORCE PTA BALLOON
Common Device Name
FOCUSED FORCE PTA BALLOON
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7684293
Report Number2020394-2018-01185
Device Sequence Number1409936
Product Code PNO
UDI-Device Identifier00801741116797
UDI-Public(01)00801741116797
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K172571
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/28/2020
Device Model NumberUS35130515
Device Catalogue NumberUS35130515
Device Lot NumberGFCP0625
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/20/2018
Initial Date Received by Manufacturer 06/14/2018
Supplement Date Received by Manufacturer09/05/2018
Initial Report FDA Received Date07/12/2018
Supplement Report FDA Received Date09/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age63 YR
Patient Weight70
-
-