| Catalog Number |
AB18W030060150 |
| Medical Device Problem Codes |
Component Missing (2306); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Image analysis: a set of four photographs of cine images was received for evaluation.Cine images one and three are identical.Cine images two and four are identical.Cine image one is of the powercross balloon inflated within the targeted vessel and the ancillary 0.018¿ nitrex guidewire.Only the powercross distal radiopaque marker band is visible.Image two is of the powercross not inflated on the 0.018¿ guidewire at the targeted vessel.Only the powercross distal radiopaque marker band is visible.Device analysis: the powercross dilatation catheter was received for evaluation.No ancillary devices from the procedure were received for evaluation.When removed from its protective transportation hoop it was noted that only the distal radiopaque marker was on the catheter.The balloon chamber was in a post inflated profile, (e.G.Not tightly wrapped or winged) and a clear fluid was noted within the balloon chamber.The proximal radiopaque marker band was found approximately 42.5cm proximal of the distal tip of the dilatation catheter.A 20cc water filled syringe was attached to the proximal hub luer lock and the guidewire lumen was flushed: a clear steady stream of fluid exited the distal tip of the dilatation catheter.A 0.018¿ guidewire was loaded through the distal tip of the dilatation catheter with ease.A 20cc water filled syringe was attached to the inflation lumen y-manifold luer lock and the balloon chamber was inflated.A vacuum was drawn, and the balloon chamber deflated.The balloon chamber was inflated to allow the examination of the balloon chamber¿s inner guidewire lumen.A witness mark was noted were the proximal radiopaque marker was swaged.Additional destructive testing was performed.The proximal radiopaque marker was located in the inflation lumen pathway.The dilatation catheter was cut proximal and distal of the radiopaque marker.The radiopaque marker slid off the inner guidewire lumen and the inner guidewire lumen was removed from the inflation pathway.The proximal radiopaque marker band did not exhibit witness marks of being swaged.There were no witness marks on the inner guidewire lumen where the marker band was located to indicate that an attempt to swage the marker band at that location.The balloon chamber was sliced opened and the inner guidewire lumen was removed from the dilatation catheter.No witness marks were noted on the inner guidewire lumen.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The physician was attempting to use balloon angioplasty treat a 100mm plaque lesion with 75% stenosis and no calcification in the mid left superficial femoral artery (sfa) using a powercross device and a nitrex 18 300cm guidewire.No tortuosity was reported.Pre-dilation was performed with 3x60 device (8atms and one inflation).After pre-dilation a 4x120 device was used (7 atm and 1 inflation).The powercross balloon usually has two radiopaque markers but it was noticed while positioning the balloon that the proximal marker on the balloon was not visible.The balloon was given opaque shape slightly to see the length of the balloon.The issue was noticed after the product was sent into the patient.The marker was not initially visible in the patient and it was reported that it did not fall off inside the patient.The procedure was completed normally.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|