| Catalog Number |
K08A |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/18/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent spinal surgery due to osteoporotic fracture.Intra-op, while inflating the balloon it bursted.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Level implanted: thoracic vertebra 12.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual functional visual and functional evaluation of the ibt balloon identified a sharp cut at the balloon, disallowing inflation.The morphology, location and timing of the balloon rupture is consistent with possible contact with sharp object, such as bone splinters.If information is provided in the future, a supplemental report will be issued.
|
| |
|