MEDTRONIC SOFAMOR DANEK USA, INC ARTHROSCOPE Back to Search Results
Catalog Number K08A
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/18/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient underwent spinal surgery due to osteoporotic fracture.Intra-op, while inflating the balloon it bursted.No patient complications were reported.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Level implanted: thoracic vertebra 12.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Visual functional visual and functional evaluation of the ibt balloon identified a sharp cut at the balloon, disallowing inflation.The morphology, location and timing of the balloon rupture is consistent with possible contact with sharp object, such as bone splinters.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTHROSCOPE
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
MDR Report Key7684739
Report Number1030489-2018-00988
Device Sequence Number920919
Product Code HRX
UDI-Device Identifier00858196001154
UDI-Public00858196001154
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K041454
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 12/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberK08A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/24/2018
Initial Date Received by Manufacturer 06/14/2018
Supplement Date Received by Manufacturer08/07/2018
08/08/2018
11/14/2018
Initial Report FDA Received Date07/13/2018
Supplement Report FDA Received Date09/04/2018
10/03/2018
12/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age74 YR
-
-