DAVIS & GECK CARIBE LTD DERMALON; SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE Back to Search Results
Model Number 88861741-21
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Information (3190)
Date of Event 06/28/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a procedure, both the needle and thread broke.The needle was detached from the suture.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a procedure, both the needle and thread broke.The needle was detached from the suture.There was no patient injury.
 
Additional Manufacturer Narrative
Post market vigilance (pmv) received one device and four boxes of representative samples sealed with the appropriate packaging.A tactile and microscopic inspection of the sutures was performed with no abnormalities.The visual inspection of the opened sample noted one suture and two needles.One needle was returned disengaged from the suture.Under microscopic inspection, normal crimp and swage marks were noted.A needle attachment test was performed on the sealed samples, and the results met the specifications for this product.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DERMALON
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key7685196
Report Number9612501-2018-01289
Device Sequence Number16370349
Product Code GAR
UDI-Device Identifier10884521070714
UDI-Public10884521070714
Combination Product (Y/N)N
Initial Reporter CountryHK
PMA/510(K) Number
K981582
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 01/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2018
Device Model Number88861741-21
Device Catalogue Number88861741-21
Device Lot NumberD3M0084X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/27/2018
Initial Date Received by Manufacturer 06/28/2018
Supplement Date Received by Manufacturer08/08/2018
10/05/2018
12/11/2018
Initial Report FDA Received Date07/13/2018
Supplement Report FDA Received Date10/03/2018
10/30/2018
01/08/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-