| Catalog Number |
03.70.157 |
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
06/15/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Batch review performed on 10 july 2018; lot 1620220: (b)(4) items manufactured and released on 13 october 2016.Expiration date:2021-09-25; no anomalies found related to the problem.To date, (b)(4)items of the same lot have been already sold without any similar reported event.
|
| |
|
Event or Problem Description
|
|
The screw broke during insertion (acdf) with drivers 03.70.0032 and 03.70.0033.Part of the screw fell into the patient, the surgeon was able to recover the broken part and fragments.A second screw was used, there was a five minutes delay.
|
| |
|
Additional Manufacturer Narrative
|
|
To date, the device was not received for investigation.No information about its effective availability.Preliminary investigation performed by r&d project manager: although no definitive root cause can be established, it is most likely that a non-standard action occurred during the device application that results in an abnormally high stress on the device.This type of breakage has not been identified as a systemic issue of the device, and the batch record review does not indicate any potential manufacturing related issue.
|
| |
|