| Model Number |
PLATINIUM CRT-D 1711 |
| Medical Device Problem Codes |
Over-Sensing (1438); Defibrillation/Stimulation Problem (1573)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/18/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Reportedly, episodes showing ventricular oversensing were observed in the device memories during a routine follow-up performed on (b)(6) 2018.In one episode, a shock was triggered but could not be delivered due to overload detection.Preliminary analysis confirmed the reported issue and showed that it was most probably due to a defibrillation lead issue.Based on available data, no issue is suspected on the icd.Recommendations to replace the defibrillation lead were sent on (b)(6) 2018.
|
| |
|
Additional Manufacturer Narrative
|
|
Please refer to the attached analysis report.
|
| |
|
Event or Problem Description
|
|
Reportedly, episodes showing ventricular oversensing were observed in the device memories during a routine follow-up performed on (b)(6) 2018.In one episode, a shock was triggered but could not be delivered due to overload detection.Preliminary analysis confirmed the reported issue and showed that it was most probably due to a defibrillation lead issue.Based on available data, no issue is suspected on the icd.Recommendations to replace the defibrillation lead were sent on (b)(6) 2018.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated ventricular lead was replaced in december 2018.
|
| |
|
Event or Problem Description
|
|
Reportedly, episodes showing ventricular oversensing were observed in the device memories during a routine follow-up performed on (b)(6)2018.In one episode, a shock was triggered but could not be delivered due to overload detection.Preliminary analysis confirmed the reported issue and showed that it was most probably due to a defibrillation lead issue.Based on available data, no issue is suspected on the icd.Recommendations to replace the defibrillation lead were sent on (b)(6)2018.
|
| |
|