• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 630G INSULIN PUMP MMT-1715K 630G BLACK MG ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 630G INSULIN PUMP MMT-1715K 630G BLACK MG ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-1715K
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/18/2017
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event. The device has been returned, but not yet evaluated. Further information will follow once the analysis has been completed. No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that the insulin was squirting out during manual prime process. The customer's blood glucose was unknown at the time of incident. The customer performed troubleshooting. The insulin did exit during process. The insulin pump will not be returned for analysis.
 
Manufacturer Narrative
Insulin pump passed the rewind test, prime or seating, basic occlusion, occlusion, force sensor test and displacement test. No prime or fill anomaly was noted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name630G INSULIN PUMP MMT-1715K 630G BLACK MG
Type of DeviceARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
8185464805
MDR Report Key7686089
MDR Text Key114041731
Report Number3004209178-2018-91348
Device Sequence Number1
Product Code OZO
UDI-Device Identifier00643169656840
UDI-Public(01)00643169656840(017)20180610
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Type of Report Initial,Followup
Report Date 09/27/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received07/13/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date06/10/2018
Device Model NumberMMT-1715K
Device Catalogue NumberMMT-1715K
Device Lot NumberHG26MXB
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/30/2018
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/30/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-