• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC COOK TPN DOUBLE LUMEN TPN CATHETER SET; FOZ CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC COOK TPN DOUBLE LUMEN TPN CATHETER SET; FOZ CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS Back to Search Results
Catalog Number C-TPNS-7.0D-65
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported a cook tpn double lumen tpn catheter set broke where the hub and silicone catheter connect.To date, additional information has been requested three times, but no additional information is available.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
 
Additional Manufacturer Narrative
A review of the complaint history, device history record, documentation, instructions for use (ifu), manufacturing instructions, specifications, trends, and quality control were conducted during the investigation.The complaint device was not returned; therefore, no physical examination could be performed.However, a document-based investigation was performed.There is no evidence to suggest the finished product was not made to specifications.A review of the device history record of the finished product shows no nonconforming events that could contribute to this failure mode.A complaint history search revealed that there were no other reported complaints for this lot number.During the investigation evaluation, a review of the current versions governing the manufacturing and quality controls was conducted.This review surmised that the current process and quality inspection checks are in place to ensure the functionally and device integrity prior to shipment.No specific issue with this documentation that may have contributed to this failure mode was found, and the complaint devices were found to be manufactured within specification.Based on the information provided, no product returned, and the results of our investigation, a definitive root cause could not be determined.Per the [quality engineering] risk assessment, no further action is required.Monitoring will continue to be performed for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK TPN DOUBLE LUMEN TPN CATHETER SET
Common Device Name
FOZ CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7686216
Report Number1820334-2018-02167
Device Sequence Number1316026
Product Code FOZ
UDI-Device Identifier00827002018808
UDI-Public(01)00827002018808(17)190906(10)7251536
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K950118
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/06/2019
Device Catalogue NumberC-TPNS-7.0D-65
Device Lot Number7251536
Was Device Available for Evaluation? No
Event Location Hospital
Initial Date Received by Manufacturer 06/26/2018
Supplement Date Received by Manufacturer09/04/2018
Initial Report FDA Received Date07/13/2018
Supplement Report FDA Received Date10/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-