COOK INC FLEXOR RAABE GUIDING SHEATH; DYB INTRODUCER, CATHETER Back to Search Results
Model Number G12242
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
During stent implantation of a lower extremity artery, it was reported the valve had detached from the flexor raabe guiding sheath.The complaint device was withdrawn from the patient's vasculature and replaced with another.The procedure was completed using the new sheath.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
The operator said the device was in good condition before use.During the procedure, the valve was detached from the near end of the sheath when the operator withdraw the balloon.The operator changed to a new sheath to continue the procedure and finished the procedure successfully.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Investigation - evaluation a review of the complaint history, device history record, dimensional verification, instruction for use, manufacture instructions, quality control, and a visual inspection of the returned device were conducted during the investigation.The visual inspection of the returned device confirmed the detached sheath from the hub.The proximal fitting was separated from the sheath tubing.The sheath tubing did not appear compressed.The device had biological matter on the side arm, proximal fittings, and in the sheath tubing.No retained tubing was visible in proximal fittings, and the flare appeared intact.The tubing was slightly bent at approximately 1.2 cm from the proximal edge of the flare.The tubing measured 55.2 cm in length.The sheath tubing did not have visible hemostat marks, scratches or kinks.Cook measured the connecting cap with a pin gauge and it was within specification.Cook gauged the flare and it was within specification.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record showed no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Furthermore, an instructions for use (ifu) is provided with the device.The ifu you states "the maximum diameter of the instrument or catheter to be introduced should be determined to ensure its passage through the introducer.All instruments or catheters used with this product should move freely through the valve and sheath.Damage to the valve/introducer may result when the fit is tight.When inserting, manipulating or withdrawing a device through an introducer always maintain introducer position.Sheath introduction upon removal from package, ensure the inner diameter (id) of the introducer is appropriate for the maximum diameter or catheter to be introduced." based on the information provided, the examination of the returned product, and the results of our investigation, a definitive root cause could not be determined.We have notified the appropriate personnel and will continue to monitor for similar complaints.Per the quality engineering risk assessment no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Additional Manufacturer Narrative
Correction: g5.This mdr is being submitted as having information not previously reported.Additional complaint investigation and record remediation was not performed.Blank fields on this form indicate the information is unknown or unavailable.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding patient and/or event details has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXOR RAABE GUIDING SHEATH
Common Device Name
DYB INTRODUCER, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7686883
Report Number1820334-2018-02025
Device Sequence Number13781913
Product Code DYB
UDI-Device Identifier00827002122420
UDI-Public(01)00827002122420(17)210315(10)8667576
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K142829
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,other
Initial Reporter Occupation Other Caregivers
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 03/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/15/2021
Device Model NumberG12242
Device Catalogue NumberKCFW-8.0-38-55-RB-RAABE
Device Lot Number8667576
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2018
Initial Date Received by Manufacturer 06/26/2018
Supplement Date Received by Manufacturer08/29/2018
10/16/2018
03/20/2020
Initial Report FDA Received Date07/13/2018
Supplement Report FDA Received Date09/26/2018
10/29/2018
03/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age67 YR
Patient Weight74
-
-