| Model Number |
11612593 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The affected product has been received and the investigation is pending.A follow up report will be submitted once the failure investigation has been completed.
|
| |
|
Event or Problem Description
|
|
The customer reported that the filter leaked during a tpn infusion.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer¿s report of a leaking filter was confirmed.Visual inspection of the set did not reveal any discernible damage or abnormalities.Functional testing confirmed a slow leak from the lower vent hole on the filter.The root cause of the leaking was not determined.
|
| |
|
Event or Problem Description
|
|
The customer reported that the filter leaked during a tpn infusion.There was no patient harm.
|
| |
|