• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION ALARIS INFUSION SET; SET, INFUSION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION ALARIS INFUSION SET; SET, INFUSION, INTRAVASCULAR Back to Search Results
Model Number 11612593
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Concomitant medical products: non-bd extension set; non-bd access bag; spiros connectors, therapy date (b)(6) 2018.The affected product has been received and the investigation is pending.A follow up report will be submitted once the failure investigation has been completed.
 
Event or Problem Description
The customer reported that the filter was in use for less than 24 hours when it leaked while infusing maintenance fluids.There was no patient harm.
 
Additional Manufacturer Narrative
Concomitant medical products: (3)spiros caps.The customer¿s report of a leaking filter was confirmed.Visual inspection of the set did not reveal any discernible damage or abnormalities.Functional testing confirmed leaking from the upper vent hole on the filter.Inspection under magnification of the leaking component did not reveal any noticeable damage to either the filter membrane or housing.The root cause of the leaking was not determined.
 
Event or Problem Description
The customer reported that the filter was in use for less than 24 hours when it leaked while infusing maintenance fluids.There was no patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS INFUSION SET
Common Device Name
SET, INFUSION, INTRAVASCULAR
Manufacturer (Section D)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
MDR Report Key7687031
Report Number9616066-2018-00993
Device Sequence Number19182057
Product Code FPA
UDI-Device Identifier10885403283963
UDI-Public10885403283963
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K970485
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 06/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number11612593
Device Catalogue Number11612593
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/13/2018
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date07/13/2018
Supplement Report FDA Received Date08/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
8100,8015, THERAPY DATE (B)(6) 2018; NON-BD BAG ACCESS DEVICE,NON-BD BI-FUSE SET
-
-