| Model Number |
CRPLUS |
| Medical Device Problem Code |
No Audible Prompt/Feedback (2282)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
06/22/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Physio-control provided the customer with technical assistance.The customer removed the device's lid and reinstalled it, which resolved the reported issue.The device was not returned to physio-control for evaluation.The cause of the reported issue could not be determined.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device did not provide any voice prompts when the device was powered on.Without voice prompts, a lay user would not receive the audible instructions on how to properly use the device on a patient.There was no patient use associated with the reported event.There was no patient use associated with the reported event.
|
| |
|