MAUDE Adverse Event Report: DEPUY MITEK LLC US INTRAFIX FEMORAL SHEATH & SCREW SYSTEM, 7.5 MM; ORTHOPAEDIC BONE SCREW, NON-BIOABSORBABLE, STERILE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US INTRAFIX FEMORAL SHEATH & SCREW SYSTEM, 7.5 MM; ORTHOPAEDIC BONE SCREW, NON-BIOABSORBABLE, STERILE Back to Search Results
Catalog Number 254681
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/05/2013
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The batch review 3449397 showed that the devices were conformed to in process and finished goods specifications when released to stock the 9 aug.2010.Expiration date: 2013-07 (according to retained labels).No nc was opened.Evaluation statement: the complaint device is not being returned, therefore is unavailable for a physical evaluation.A batch record review has been conducted to determine if there were any internal processing issues which would have contributed to the nature of the product complaint.Our results indicate that this batch of product was processed without incident and therefore there is no internally assignable cause for the reported problem.Further, a review into the depuy mitek complaints system revealed 1 other complaint for this lot of devices that were released to distribution, but the failure mode was for a screw cracking and not a full breakage.These complaints are not related to each other based of failure mode.We cannot discern a root cause for the reported failure mode.At this point in time, no corrective action is required and no further action is warranted.However, depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.This report is being filed from the remetrex complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The sales rep reported one of his intrafix 7.5mm femoral screws broke upon insertion into the sheath during an acl procedure.He stated the device started to crack as the surgeon inserted the device and continued anyways and then the screw broke in two pieces.The surgeon removed all fragments of the screw but kept the sheath in the patient then inserted a new screw to complete the procedure with no patient consequences or delays.The device was already discarded by the customer.The sales rep stated this was a (b)(6) female with hard bone and believes the surgeon should have used a different type of screw for the type of bone.No further details are available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRAFIX FEMORAL SHEATH & SCREW SYSTEM, 7.5 MM
Common Device Name
ORTHOPAEDIC BONE SCREW, NON-BIOABSORBABLE, STERILE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key7687232
Report Number1221934-2018-51145
Device Sequence Number683665
Product Code MAI
UDI-Device Identifier10886705013623
UDI-Public10886705013623
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
K150209
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/05/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2013
Device Catalogue Number254681
Device Lot Number3449397
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/05/2013
Initial Report FDA Received Date07/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age19 YR
-
-