Model Number CRPLUS
Medical Device Problem Code Unintended Power Up (1162)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 06/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Physio-control continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
 
Event or Problem Description
The customer contacted physio-control to report that their device was powering itself on without any user interaction and it would have to be manually powered off.As a result, the device's internal batteries may become prematurely depleted and the device may not have enough power to provide adequate defibrillation therapy if needed.There was no patient use associated with the reported event.
 
Additional Manufacturer Narrative
Physio-control sent the customer a replacement charge-pak.The customer installed the replacement charge-pak in their device and confirmed that it resolved the reported issue.The device was not returned to physio-control for evaluation.The cause of the reported issue could not be determined.
 
Event or Problem Description
The customer contacted physio-control to report that their device was powering itself on without any user interaction and it would have to be manually powered off.As a result, the device's internal batteries may become prematurely depleted and the device may not have enough power to provide adequate defibrillation therapy if needed.There was no patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK CR® PLUS DEFIBRILLATOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key7687357
Report Number0003015876-2018-01106
Device Sequence Number1315605
Product Code MKJ
UDI-Device Identifier00885074440323
UDI-Public00885074440323
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K033275
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberCRPLUS
Device Catalogue Number99403-000156
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 06/22/2018
Supplement Date Received by Manufacturer07/31/2018
Initial Report FDA Received Date07/13/2018
Supplement Report FDA Received Date08/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-