• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION ALARIS INFUSION SET; SET, INFUSION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION ALARIS INFUSION SET; SET, INFUSION, INTRAVASCULAR Back to Search Results
Model Number 11612593
Medical Device Problem Codes Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/04/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The affected product has been received and the investigation is pending.A follow up report will be submitted once the failure investigation has been completed.
 
Event or Problem Description
The customer reported that the tubing set had a hole, as well as, the filter was leaking during patient use.There was no patient harm.
 
Additional Manufacturer Narrative
The customer¿s report of a hole in the tubing was confirmed.The report of a leaking filter was not confirmed.Examination under magnification showed 2 short, rough-edged cuts approximately 8¿ below the lower fitment.The 2 cuts together formed a v shape, similar to what would be observed after forceps or another clamping instrument had been used.Leaks were noted from the 2 holes, but no leaking from the filter was observed.The cause of the damage and leaking was determined to be the use of forceps or a similar clamping instrument.
 
Event or Problem Description
The customer reported that the tubing set had a hole, as well as, the filter was leaking during patient use.There was no patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS INFUSION SET
Common Device Name
SET, INFUSION, INTRAVASCULAR
Manufacturer (Section D)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
MDR Report Key7687399
Report Number9616066-2018-00995
Device Sequence Number16302268
Product Code FPA
UDI-Device Identifier10885403283963
UDI-Public10885403283963
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K970485
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 06/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number11612593
Device Catalogue Number11612593
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/13/2018
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date07/13/2018
Supplement Report FDA Received Date08/16/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-