| Model Number |
11612593 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/04/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The affected product has been received and the investigation is pending.A follow up report will be submitted once the failure investigation has been completed.
|
| |
|
Event or Problem Description
|
|
The customer reported that the tubing set had a hole, as well as, the filter was leaking during patient use.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer¿s report of a hole in the tubing was confirmed.The report of a leaking filter was not confirmed.Examination under magnification showed 2 short, rough-edged cuts approximately 8¿ below the lower fitment.The 2 cuts together formed a v shape, similar to what would be observed after forceps or another clamping instrument had been used.Leaks were noted from the 2 holes, but no leaking from the filter was observed.The cause of the damage and leaking was determined to be the use of forceps or a similar clamping instrument.
|
| |
|
Event or Problem Description
|
|
The customer reported that the tubing set had a hole, as well as, the filter was leaking during patient use.There was no patient harm.
|
| |
|