| Catalog Number |
70-6529 |
| Medical Device Problem Codes |
Failure to Sense (1559); Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Corrosion of the analyzer sensor pins were noted during evaluation of the analyzer, which is the most likely cause of the screen malfunction, error messages and failed qc results.Corrosion of the sensor pins occurs when there is over exposure of the analyzer pins to the assay's acidic treatment reagent.This occurs when used test sensors are not removed immediately after testing or when too much sample is added to the test sensor.Per sections 4-8 and 5-8 of the leadcare ii user's manual, "remove the used sensor immediately after recording the result." investigation of the event log showed the customer had been experiencing repeated run-time error messages prior to contacting magellan.(b)(4).
|
| |
|
Event or Problem Description
|
|
A user facility called magellan's product support on (b)(6) 2018 to report that the message screen of their leadcare blood lead analyzer was not advancing properly from test step to test step.There was no indication at the time of the report that the analyzer was reporting erroneous results.The analyzer was returned to magellan for evaluation on (b)(6) 2018.During investigative testing, magellan observed repeated failed external control tests.A review of the analyzer's event logs showed that the instrument had generated numerous error messages over several months while in use at the user facility but these had not been reported to magellan as specified in product labeling.In addition, the event logs indicate that external qc was not consistently being performed on a monthly basis in accordance with label instructions.There have been no reports of injuries or erroneous results reported by the user facility.
|
| |
|
Search Alerts/Recalls
|