| Catalog Number |
TGU404020 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Codes |
Vascular Dissection (3160); Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
06/15/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The review of the manufacturing paperwork verified that this lot met all pre-release specifications.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, the patient underwent endovascular repair of a thoracic aortic dissection using a conformable gore® tag® thoracic endoprosthesis.It was reported that after the gore® tag® thoracic endoprosthesis (tgu404020/17752645) was deployed, upon retracting the delivery catheter the leading olive met resistance on the sheath.The physician was unaware that the olive tip had separated and advanced a catheter on the guidewire and pushed the olive off the guidewire into the patient's left common iliac artery.The physician chose to deploy a stent and capture the olive between the left common iliac wall and the stent and leave it in the patient.The patient tolerated the procedure.
|
| |
|