• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W.L. GORE & ASSOCIATES AORTIC TAG THORACIC ENDOPROSTHESIS (CTAG); SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W.L. GORE & ASSOCIATES AORTIC TAG THORACIC ENDOPROSTHESIS (CTAG); SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Catalog Number TGU404020
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Codes Vascular Dissection (3160); Device Embedded In Tissue or Plaque (3165)
Date of Event 06/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The review of the manufacturing paperwork verified that this lot met all pre-release specifications.
 
Event or Problem Description
On (b)(6) 2018, the patient underwent endovascular repair of a thoracic aortic dissection using a conformable gore® tag® thoracic endoprosthesis.It was reported that after the gore® tag® thoracic endoprosthesis (tgu404020/17752645) was deployed, upon retracting the delivery catheter the leading olive met resistance on the sheath.The physician was unaware that the olive tip had separated and advanced a catheter on the guidewire and pushed the olive off the guidewire into the patient's left common iliac artery.The physician chose to deploy a stent and capture the olive between the left common iliac wall and the stent and leave it in the patient.The patient tolerated the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AORTIC TAG THORACIC ENDOPROSTHESIS (CTAG)
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W.L. GORE & ASSOCIATES
flagstaff AZ
MDR Report Key7687787
Report Number2017233-2018-00377
Device Sequence Number1315606
Product Code MIH
UDI-Device Identifier00733132617760
UDI-Public00733132617760
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P040043
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/12/2021
Device Catalogue NumberTGU404020
Device Lot Number17752645
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date07/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age49 YR
-
-