| Model Number |
420189-09 |
| Medical Device Problem Codes |
Break (1069); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/16/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) received the instrument involved with this complaint and completed the device evaluation.Failure analysis confirmed the customer reported failure.The instrument was found to have a broken pitch cable at the distal clevis hub.The broken cable segment that contains the crimp was missing from clevis.Isi has conducted a device history record (dhr) review for this device and did not find any non-conformances that were related to this reported event.The customer reported complaint does not itself constitute an mdr reportable event; however, the broken pitch cable found during failure analysis could cause or contribute to an adverse event if the malfunction were to recur.
|
| |
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted pulmonary wedge resection procedure, the double fenestrated grasper had a broken cable.There was no report of fragment(s) falling inside the patient, patient harm, adverse outcome or injury.
|
| |
|
Search Alerts/Recalls
|