| Model Number |
B24802 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/15/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A field service engineer (fse) evaluated the instrument and did not find an instrument malfunction.The cause of the false high plt results is unknown.The bec internal identifier for this event is case-(b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reported false high platelet (plt) results were generated on a unicel dxh 800 coulter cellular analysis system.There were no flags or system messages reported by the customer.The customer reran the sample on a competitive analyzer, which generated lower results.The competitive analyzer results correlated with the plt count from a manual slide review.There was no effect or change to patient treatment associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Follow up 01: an improvement for plt flagging was developed and released as a part of a software update to the dxh800 that is designated v3.2.1 on (b)(6) 2017; approximately 11.5% of the worldwide product installed base has been upgraded as of (b)(6) 2018.Due to an increase in the number of complaints received related to unflagged erroneously high plt patient results released from the lab, bec has determined that a field action is warranted.Field action (fa-33718) was approved on 30-jul-2018.The field action includes a software patch for the improvement for the plt flagging component of v3.2.1, which will expedite installation of this improvement for rest of the installed base.Section h2 - additional information has been selected.Section h7 - recall, notification and modification have been selected section h9 - number has been updated to reflect the respective information for this event.Additional information: the recall (fa-33718) includes notification to the customer and correction of the issue via software update.Bec internal identifier - case(b)(4).
|
| |
|
Search Alerts/Recalls
|