SMITH & NEPHEW, INC. LGN OFFSET COUPLER TRIAL 6MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS
|
Back to Search Results |
|
| Catalog Number |
71434306 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
06/25/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Dr needed to remove the trial, using the slap hammer, dr realized the femoral component came away from the bone, without the stem and offset coupler.On investigation we realized the coupler had broken, part in the femoral trial and part in the trial stem in the medullary canal.He needed to remove some bone from the anterior cortex to gain access to the end of the construct to enable him to grip it with bone nibblers and remove it.On further examination we realized the locking screw in the coupler had fractured.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated device was returned and evaluated.The clinical/medical team concluded, no clinically supporting documents have been provided for inclusion in this investigation.However, it was reported no patient injury occurred as a result of the instrument breakage intra-operatively.Based on the information provided, no further clinical/medical assessment is warranted at this time.A visual examination of the returned legion offset coupler trial indicated that it has fractured in half.The connection screw is assembled through the offset coupler male part into the offset coupler female part with a thread locking agent.An additional set screw is assembled on top of the connection screw in the offset coupler male part to fasten it during surgery.If an excessive amount of torque is applied to the set screw, torque may be transferred to the connection screw leading to a fracture.The cause for the break in the offset coupler trial connection screw may be attributed to excessive torque applied to the set screw.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate further as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.
|
| |
|
Search Alerts/Recalls
|
|
|