| Model Number |
MMT-523RNAS |
| Medical Device Problem Code |
Device Difficult to Program or Calibrate (1496)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/18/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump alarmed button error.Customer's blood glucose level was 168 mg/dl.The insulin pump was exposed to moisture and had a blank display.Customer was advised to discontinue use of the device and revert to a backup plan.The customer was advised that the device would be replaced and agreed to return the product for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with blank display due to moisture damaged electronic assembly.Unable to perform operating current test, unexpected restart error test, self test and displacement test or verify button error alarm, keypad anomaly due to blank display.Device had corrode esc button during visual inspection.
|
| |
|