| Model Number |
482400J |
| Medical Device Problem Code |
No Audible Alarm (1019)
|
| Health Effect - Clinical Codes |
No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported to covidien that the customer had an issue with the enteral feeding pump.On (b)(6) 2018, the unit was triaged and service found that the pump failed the buzzer test.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation of the kangaroo pump was performed and the customer states, ¿ pump was not working." the unit was triaged on (b)(6) 2018, service tech confirmed customer's reported condition and found that the pump failed the buzzer test (no audible alarm).A trend has been identified and a correction action has been opened to address this issue.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|