• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. ELLIPSE VR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. ELLIPSE VR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD1275-36Q
Medical Device Problem Code Fitting Problem (2183)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/17/2018
Type of Reportable Event Malfunction
Event or Problem Description
Related to manufacturer report number: 2938836-2018-08519.During follow-up, increased high voltage lead impedance was observed on the right ventricular (rv) lead.A new rv lead was implanted to try and resolve the issue but the impedance remained high.The issue was resolved by explanting and replacing the device.The old lead was connected and the new lead was electively explanted.The patient was in stable condition and experienced no adverse health consequences.
 
Additional Manufacturer Narrative
Additional information: visual inspection revealed no anomalies.Resistance measurements were performed to verify the header to hybrid connections.High impedance was found on the right ventricular (rv) header to hybrid connection.Diagnostic imaging revealed the rv connector block wire to be broken at the weld.The reported event of increased high voltage (hv) lead impedance was confirmed.The broken connection caused the increased hv lead impedance observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELLIPSE VR
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
645 almanor avenue
sunnyvale CA 94085
MDR Report Key7690019
Report Number2938836-2018-08522
Device Sequence Number16345061
Product Code NVZ
UDI-Device Identifier05414734506038
UDI-Public05414734506038
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P910023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2018
Device Model NumberCD1275-36Q
Device Lot NumberS000024522
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/21/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/21/2018
Supplement Date Received by Manufacturer10/12/2018
Initial Report FDA Received Date07/16/2018
Supplement Report FDA Received Date10/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
(B)(4)
-
-