| Model Number |
01-207-0007 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Electric Shock (2554)
|
| Date of Event |
06/18/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Complaint received that alleges "when 1st got the device there was no issues using it, but on (b)(6) 2018 she got a shock while using the device & was in pain.Said she has gone to the emergency room (er) & was informed that she has titanium rods implanted & so the magnetic waves was causing the shocks to her body.Said she also talked to her dr & dr has informed her to stop using the device.Patient states has not used the device since (b)(6) 2018 but is still getting shocks inside her body".
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for evaluation.Full treatment was ran at a test station and observed results were within specifications.A quality assurance technician wore the device for a full 30 minute treatment with no shocks, pain, or discomfort during or after treatment.The reported event could not be duplicated.The device performed as expected with no failure identified.
|
| |
|