Event or Problem Description
|
|
Complainant alleged that while attempting to treat a (b)(6) patient (gender unknown) for cardiac arrest, the device failed to discharge multiple times.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation and the customer's report was not replicated or confirmed.The device was put through extensive testing without duplicating the report.The device was recertified and returned to the customer.The electrode pads were not available for evaluation as part of this investigation.No clinical file was available.No trend is associated with reports of this type.
|
| |
|