| Catalog Number |
8470107030DC |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
06/20/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A follow up report will be filed once the quality investigation is complete.
|
| |
|
Event or Problem Description
|
|
It was reported that during service conducted at the manufacturer a broken bur was found inside the attachment.It was also reported that this event did not occur during a surgical procedure and there was no delay or adverse consequences as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results indicate that the device broke as a result of a bearing failure in the associated attachment (5407120970c (b)(4)).
|
| |
|
Event or Problem Description
|
|
It was reported that during service conducted at the manufacturer a broken bur was found inside the attachment.It was also reported that this event did not occur during a surgical procedure and there was no delay or adverse consequences as a result of this event.
|
| |
|
Search Alerts/Recalls
|