| Catalog Number |
HBD-15-16.5-18 |
| Medical Device Problem Codes |
Leak/Splash (1354); Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/13/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical product: medtronic everest 30 inflation device.Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: the additional information provided from the user indicated the instructions for use were not followed in regard to setup and inflation of the device.This is the most likely cause for the reported observation.Specifically the user indicated the following: 1.Negative pressure was not applied to the balloon prior to advancement through the endoscope accessory channel.The instructions for use advise the user: "to facilitate passage through the endoscope, apply negative pressure to the catheter." the application of negative pressure will aspirate all residual air from the balloon and ease endoscopic advancement.Negative pressure will also aid in balloon preservation and optimize balloon performance.The instructions for use advise the user: "maintain balloon deflation with negative pressure and introduce into the accessory channel of the endoscope, advancing in short increments until the dilator is completely visualized endoscopically." 2.The balloon was inflated using a medtronic everest 30 inflation device.The medtronic everest 30 inflation device has a 20 cc syringe which is incompatible with the dilation balloon and cannot be used to inflate it.The instructions for use advise the user: "attach the balloon to a 60 ml (cc) inflation device with gauge to monitor balloon pressure." 3.Contrast was used to inflate the balloon.The instructions for use advise the user: "the balloon must be inflated with water only." prior to distribution, all hercules 3 stage balloons esophageal are subjected to a visual examination to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted and this represents an unusual occurrence.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.Additional comments regarding this report: based on the information provided that negative pressure was not applied, the balloon was inflated using an incompatible inflation device, contrast was used to inflate the balloon and that the device was used on a pediatric patient, a cook representative has been directed to contact the medical facility involved in an effort to promote further education and understanding related to appropriate usage of this product.
|
| |
|
Event or Problem Description
|
|
During a dilation procedure, the physician used a cook hercules 3 stage balloon esophageal.The balloon of the catheter pierced during the procedure.When the doctor pushed the balloon into the esophagus, the nurse inflated the balloon.The balloon did not inflate.It was punctured when she increased the pressure.There was contrast in the esophagus.A second balloon [of the] same reference was used without any incident.There were no consequences for the patient except a delay in care and ten (10) extra minutes under anesthesia.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: medtronic everest 30 inflation device.Investigation evaluation: our laboratory evaluation of the product said to be involved confirmed the report.During a functional test, a cook ds-60cc-s dilation syringe was filled with water and attached to the balloon inflation port.The syringe was placed into an inflation handle, and negative pressure was applied to the balloon.After applying negative pressure, inflation of the balloon was attempted.The balloon would not hold pressure and leakage was observed from two pinholes in the middle of the balloon material.A visual examination of the catheter showed no kinks or bends.A product-specific discrepancy that could have caused or contributed to this observation was not observed during our laboratory analysis.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: the additional information provided from the user indicated the instructions for use were not followed in regard to setup and inflation of the device.This is the most likely cause for the reported observation.Specifically the user indicated the following: negative pressure was not applied to the balloon prior to advancement through the endoscope accessory channel.The instructions for use advise the user: "to facilitate passage through the endoscope, apply negative pressure to the catheter." the application of negative pressure will aspirate all residual air from the balloon and ease endoscopic advancement.Negative pressure will also aid in balloon preservation and optimize balloon performance.The instructions for use advise the user: "maintain balloon deflation with negative pressure and introduce into the accessory channel of the endoscope, advancing in short increments until the dilator is completely visualized endoscopically." the balloon was inflated using a medtronic everest 30 inflation device.The medtronic everest 30 inflation device has a 20 cc syringe which is incompatible with the dilation balloon and cannot be used to inflate it.The instructions for use advise the user: "attach the balloon to a 60 ml (cc) inflation device with gauge to monitor balloon pressure." contrast was used to inflate the balloon.The instructions for use advise the user: "the balloon must be inflated with water only." prior to distribution, all hercules 3 stage balloons esophageal are subjected to a visual examination to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted and this represents an unusual occurrence.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.Additional comments regarding this report: based on the information provided that negative pressure was not applied, the balloon was inflated using an incompatible inflation device, contrast was used to inflate the balloon and that the device was used on a pediatric patient, a cook representative has been directed to contact the medical facility involved in an effort to promote further education and understanding related to appropriate usage of this product.
|
| |
|
Event or Problem Description
|
|
During a dilation procedure, the physician used a cook hercules 3 stage balloon esophageal.The balloon of the catheter pierced during the procedure.When the doctor pushed the balloon into the esophagus, the nurse inflated the balloon.The balloon did not inflate.It was punctured when she increased the pressure.There was contrast in the esophagus.A second balloon [of the] same reference was used without any incident.There were no consequences for the patient except a delay in care and ten (10) extra minutes under anesthesia.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
| |
|