| Model Number |
CRPLUS |
| Medical Device Problem Code |
Power Problem (3010)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
07/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The customer was provided with a replacement device.Physio-control continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device displayed all three icons (charge-pak, attention, and service wrench).The illumination of all three icons indicates that the device may not be able to provide sufficient defibrillation therapy, if necessary.There was no patient use reported for this event.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device displayed all three icons (charge-pak, attention, and service wrench).The illumination of all three icons indicates that the device may not be able to provide sufficient defibrillation therapy, if necessary.There was no patient use reported for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Physio-control evaluated the customer¿s device and was unable to duplicate the reported issue.The device was able to power on as normal and no discrepancies were observed on the readiness indicator of the device.The cause of the reported issue could not be conclusively determined.The customer was sent a replacement device.
|
| |
|