(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the complaint device was received and evaluated.Visual observation reveals the color of the handle is slightly faded indicating it is old and seen heavy use.The trigger handle is loose and shaking as if it was missing an inner component that holds it in place.To test its functionality, a versalok ti anchor was loaded and locked.When the trigger handle is actuated, the pin does not deploy into the sleeve of the anchor.The failure of this device to perform its intended operation is confirmed.From the visual appearance, it has been determined that this device was excessively abused beyond its design intent and its failure could be attributed to field wear.A batch record review has been conducted to determine if there were any internal processing issues which would have contributed to the nature of the product complaint.Our results indicate that this batch of product was processed without incident and therefore there is no internally assignable cause for the reported problem.Further, a review into the depuy mitek complaints system revealed no other complaints of any kind for this lot of devices that were released to distribution.This device is 2 years old and has seen heavy use in the field and therefore, at this point in time, no further action is warranted.However, depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).Associated medwatch: [mr#1221934-2018-51310]; [mr#1221934-2018-51311].This report is being filed from the remetrex complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
|
|
Our affiliate reported the gun was in the locked position, however, the surgeon was not able to deploy the anchor pin into the sleeve by squeezing the trigger.He also tested the gun again by using another anchor with the same result.The surgeon finished the procedure with a pushlock anchor.No patient harm reported.Addit info the device is being discarded.
|