| Model Number |
MMT-751PNAH |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Hypoglycemia (1912)
|
| Date of Event |
06/18/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
Customer reported via phone call that they experienced low blood glucose level.Customer's blood glucose value was 45 mg/dl at the time of the incident.Customer will not returned device for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with blank display due to moisture damage at electronic assembly.Also, moisture damage motor during visual inspection.Unable to perform operating currents, self test, rewind test and displacement test due to blank display.(b)(4).
|
| |
|