| Model Number |
QC-8-20-3D |
| Medical Device Problem Code |
Separation Failure (2547)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/10/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received a report that the healthcare professional (hcp) attempted to detach the coil with an instant detacher but it would not detach.A manual detachment was then attempted but the coil remained undetached.As a result the coil was removed from the patient and a new device used to complete the surgery.The patient was undergoing embolization surgery to treat a 14mm cerebral aneurysm.As a result of the issue the procedure was extended less than 30 minutes.It was noted that there were 3 coils implanted using the same microcatheter prior to the issue and no abnormalities were seen.There were no related patient symptoms.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|