• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US HEALIX BR ANCHOR, 5.5 MM - W/ORTHOCORD & NEEDLES; SOFT-TISSUE ANCHOR, BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US HEALIX BR ANCHOR, 5.5 MM - W/ORTHOCORD & NEEDLES; SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Catalog Number 222232
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/07/2013
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The complaint devices were received and evaluated.Visual observation reveals that the received anchor is broken transversally along the cancellous threads.This complaint can be confirmed anchor breakages are typically associated with off axis insertion, levering during insertion, or using incorrect instrumentation for preparing bone hole.The ifu instructs using an awl-tap for this anchor.Using only an awl in a good/hard bone could lead to an increased risk of anchor breakage while inserting.There was no information reported on the tools used to prepare the bone.A batch review was conducted to determine if there were any internal processing issues, which would have contributed to the nature of the product complaint.Our results indicate that this batch was processed without incident and therefore there is no internally assignable cause for the reported problem.Further, a review into the mitek complaints system revealed no other complaints of any type for this lot of devices that were released to distribution.We cannot discern a root cause for this failure.Based on the overall complaint rate for this failure, at this point in time, no further action is warranted.However, mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.This report is being filed from the remetrex complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
 
Event or Problem Description
It was reported by affiliate that the anchor broke when it was being introduced into the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEALIX BR ANCHOR, 5.5 MM - W/ORTHOCORD & NEEDLES
Common Device Name
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key7693795
Report Number1221934-2018-51345
Device Sequence Number1108280
Product Code MAI
UDI-Device Identifier10886705007783
UDI-Public10886705007783
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K150209
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/10/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/23/2016
Device Catalogue Number222232
Device Lot Number3687964
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2014
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/10/2013
Initial Report FDA Received Date07/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age39 YR
-
-