| Catalog Number |
1647G |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested and the following was obtained: what is meant by 'suture was took off'?- did the needle pull off/did the suture break?/ did the needle break/other? explain the non-conformance/defect noted with the devices? need pull off.Specific number of devices that presented with defect in this procedure.4ea.When was the defect noted or the issue occur with each involved device- while in the package/during removal from the package/during actual use on patient (suturing)/ other? during actual use on patient.If multiple procedures were involved and were not previously reported, additional files are to be opened one file for each procedure proving the following information: on single procedure.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a surgical procedure on an unknown date and suture was used.During the procedure, the needle pulled off the suture.No adverse patient consequences were reported.Additional information had been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Pc-(b)(4).Date sent to the fda: (b)(4) 2018.It was reported pull off suture needle.An opened box with six unopened samples were returned for analysis.During the visual inspection of unopened samples, no defects were found on the packages.The samples were opened and the swage and attachment area of the needles were as expected.The sutures were dispensed without problems and examined along of the strands and no defects were observed.A functional test was performed and the pull force were above the minimum requirements.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this batch.Per the conditions of the samples received, no attachment defects were found and the tested samples met the finished goods requirements.
|
| |
|
Search Alerts/Recalls
|