BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA. BD ANGIOCATH¿ IV CATHETER; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 381147
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/22/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
There were multiple lot numbers reported to be involved.The information for each lot number is as follows: medical device lot #: 7173832; medical device expiration date: 2022-06-30; device manufacture date: 2017-07-12; medical device lot #: unknown; medical device expiration date: unknown; device manufacture date: unknown.A sample is available for evaluation.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported with the use of the bd angiocath¿ iv catheter there was an issue with foreign matter.There was no report of injury or further medical intervention.
 
Additional Manufacturer Narrative
Due to an it issue beginning on (b)(6) 2018, previously filed emdrs did not contain required fields.This supplemental emdr is filed to provide the following omitted fields: event attributed to: other.Device single use?: no.Device returned to manufacture: yes.
 
Event or Problem Description
It was reported with the use of the bd angiocath iv catheter there was an issue with foreign matter.There was no report of injury or further medical intervention.
 
Additional Manufacturer Narrative
One sample from each of the reported lot numbers was returned to flks for ftir analysis.As per request, all returned samples were examined an all exhibited clear droplets of material on the surface of the catheter.A small portion of this material was removed from a representative sample and prepared for ftir spectral analysis.The spectral analysis shows that this material is silicone.A lot history review was carried out and no related non conformances were raised in association with this packaged lot concluding all inspections were performed per the applicable operations and met qc specifications.Investigation conclusion: based on the investigations conducted for complaints of excess silicone of angiocath, it has been found that the root cause of this on the catheter is caused by the excessive amount of silicone that is dispensed during the catheter siliconization process.
 
Event or Problem Description
It was reported with the use of the bd angiocath iv catheter there was an issue with foreign matter.There was no report of injury or further medical intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ANGIOCATH¿ IV CATHETER
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA.
ave. pres.juscelino kubitschek
273 francisco bernardino
juiz de fora
MDR Report Key7694721
Report Number9610048-2018-00094
Device Sequence Number13636472
Product Code FOZ
UDI-Device Identifier00382903811472
UDI-Public00382903811472
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K151698
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,other,user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number381147
Device Lot NumberSEE H.10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/05/2018
Initial Date Received by Manufacturer 06/22/2018
Supplement Date Received by Manufacturer06/22/2018
06/22/2018
Initial Report FDA Received Date07/17/2018
Supplement Report FDA Received Date09/13/2018
10/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-