MAUDE Adverse Event Report: APOLLO ENDOSURGERY, INC LAP-BAND SYSTEM 10CM; ADJUSTABLE GASTRIC BAND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APOLLO ENDOSURGERY, INC LAP-BAND SYSTEM 10CM; ADJUSTABLE GASTRIC BAND Back to Search Results
Model Number B-2220
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Inflammation (1932); Irritation (1941); Pain (1994)
Date of Event 04/10/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Taper ii.Medwatch sent to the fda on 17/jul/2018.Device evaluation summary: the device was returned to apollo for analysis, and visual examination confirmed the connector type as taper ii.Needle marks were noted on the port septum and scratches were observed on the port housing.The band ring and shell were noted to be separated near the band buckle.An air leak test was performed and the band was noted to be leaking from the shell where the band had been separated near the buckle.A fill inspection test was performed and no blockage was observed.Under microscopic analysis, both ends of the separation of the band ring and shell, located approximately 2.85 inches from the band belt, were noted to have striated edges, consistent with damage from a surgical end cut to remove the device.Both ends of the partial separation of the band buckle and both ends of the band tubing were noted to have striated edges, consistent with damage from a surgical tool.Material degradation was noted to be covering approximately 70% of the outer surface of the band ring.Device labeling addresses the reported event as follows: adverse events: it is important to discuss all possible complications and adverse events with your patient.Complications which may result from the use of this product include the risks associated with the medications and methods utilized in the surgical procedure, the risks associated with any surgical procedure and the patient's degree of intolerance to any foreign object implanted in the body.Ulceration, gastritis, gastroesophageal reflux, heartburn, gas bloat, dysphagia, dehydration, constipation, and weight regain have been reported after gastric restriction procedures.Other adverse events considered related to the lap-band system that occurred in fewer than 1% of subjects included: esophagitis, gastritis, hiatal hernia, pancreatitis, abdominal pain, hernia, incisional infection, infection, redundant skin, dehydration, gi perforation, diarrhea, abnormal stools, constipation, flatulence, dyspepsia, eructation, cardiospasm, hematemesis, asthenia, fever, chest pain, incision pain, contact dermatitis, abnormal healing, edema, paresthesia, dysmenorrhea, hypochromic anemia, band leak, cholecystitis, esophageal dysmotility, esophageal ulcer, esophagitis, port displacement, port site pain, spleen injury and wound infection.Warnings: patients should be advised that the lap-band ap system is a long-term implant.Explant (removal) and replacement surgery may be indicated at any time.Medical management of adverse reactions may include explantation.Revision surgery for explantation and replacement may also be indicated to achieve patient satisfaction.
 
Event or Problem Description
Reported as: a patient with the lap-band system was reported to have "abdominal pain issues gastric polyp and mild distal esophagitis." the device was removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAP-BAND SYSTEM 10CM
Common Device Name
ADJUSTABLE GASTRIC BAND
Manufacturer (Section D)
APOLLO ENDOSURGERY, INC
1120 s. capital of texas hwy
bldg. 1, ste. 300
austin TX 78746
MDR Report Key7694764
Report Number3006722112-2018-00164
Device Sequence Number11080001
Product Code LTI
UDI-Device Identifier10811955020169
UDI-Public10811955020169
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P000008
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2005
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device No Information
Device Expiration Date04/01/2007
Device Model NumberB-2220
Device Catalogue NumberB-2220
Device Lot Number1091332
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/05/2018
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/18/2018
Initial Report FDA Received Date07/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age53 YR
Patient Weight107
-
-