• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD ANGIOCATH IV CATHETER; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD ANGIOCATH IV CATHETER; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 381157
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/22/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Multiple lot numbers: there were multiple lot numbers reported to be involved.The information for each lot number is as follows: medical device lot #: 7209534; medical device lot #: unknown; medical device expiration date: 07/31/2022; medical device expiration date: unknown; device manufacture date: 09/05/2017; device manufacture date: unknown.A sample is available for evaluation.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the bd angiocath iv catheter had sticky foreign matter.There was no report of exposure, serious injury or medical intervention.
 
Additional Manufacturer Narrative
Correction: due to an it issue beginning on 7/3/2018, previously filed emdrs did not contain required fields.This supplemental emdr is filed to provide the following omitted fields: event attributed to: other.Device single use?: no.Device returned to manufacture: yes.Investigation summary: "the device history report for lot number 7209534, 4325589, 4085371 and 3172184 were reviewed, no related abnormalities were found.Material number 381157 for lot 7209534, 4325589, 4085371 and 3172184 were manufactured on aug-31-2017, dec-19-2014, aug-05-2014 and jul-12-2013 respectively.This is the first instance of this failure mode with this lot number.According to the sampling plan applied for product performance, this lot was accepted and released, with no defects being noted during final assembly or in packaging visual inspections.A sample was provided for the purpose of aiding our quality engineer's investigation, inspection of the device found excessive silicon on the devices.The most probable root cause for this issue is related to the manufacturing process, silicon lubricant is applied to the catheter.Bd engineers have initiated corrective actions in response to this issue, and notified the appropriate personnel.Bd will continue to track and trend for this issue.
 
Event or Problem Description
It was reported that the bd angiocath iv catheter had sticky foreign matter.There was no report of exposure, serious injury or medical intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ANGIOCATH IV CATHETER
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MDR Report Key7694948
Report Number9610847-2018-00224
Device Sequence Number920992
Product Code FOZ
UDI-Device Identifier30382903811572
UDI-Public30382903811572
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K151698
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,other,user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number381157
Device Lot NumberSEE H.10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/05/2018
Initial Date Received by Manufacturer 06/22/2018
Supplement Date Received by Manufacturer06/22/2018
Initial Report FDA Received Date07/17/2018
Supplement Report FDA Received Date09/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-