| Catalog Number |
210800 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Codes |
Tissue Damage (2104); Foreign Body In Patient (2687)
|
| Date of Event |
06/11/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reports that the post surgery radiography of the rotator cuff repair with the double row technique ( under locoregional anesthesia ) showed a grubbing-up of the anchor which was localised in the subacromial space.During the surgery, the anchor on the lateral side was impacted and locked with a gun.The surgical suture of the cuff was removed.A new surgery occured the following morning under general anesthesia with a simple row technique.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the complaint device was received and inspected.The complaint can be confirmed.The anchor was received with the cut orthocords still attached.By visual inspection, the anchor appeared to be free of any gross visual defects.Per the ifu, depuy mitek anchors are designed to lock into cortical or cancellous bone.Bone stock must be adequate to allow proper and secure anchor placement.Incomplete insertion, overtensioning or poor bone quality may result in anchor pullout.It is possible that the surgeon did not adequately tap the anchor into the bone hole, or adequately prepare the bone hole for insertion.However, given the information provided we cannot discern a definitive root cause for the reported failure.A non-conformance search was conducted to investigate any defects during production identified that may contribute to the complaint condition.No non-conformance was identified for this part-lot number combination.At this point, no corrective action is required and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device history lot: no non-conformance was identified for this part/lot number combination.Udi: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate udi: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reports that the post surgery radiography of the rotator cuff repair with the double row technique (under locoregional anesthesia) showed a grubbing-up of the anchor which was localised in the subacromial space.During the surgery, the anchor on the lateral side was impacted and locked with a gun.The surgical suture of the cuff was removed.A new surgery occured the following morning under general anesthesia with a simple row technique.Additional information from affiliate: there was no delay.A revision occured the next day (locoregional anesthesia, general anesthesia) : 30-40 minutes, without problem, but the surgical technique was changed, no modification of the post-op instructions.The constraints for the patient : feeling of failure, a new hospitalization, a new anesthesia so a high risk.No risk increased.
|
| |
|